MIL-M-38510/121A
5. (Cont):
Increase the vertical plug-in gain to 1 mV/cm and the horizontal gain to 1 μs/cm. Toggle 81 to ground and observe the shift in the amplifier output. This is the value of 1 L8B and should be approximately 2.4 mV. With 81 in position
2, observe the amplifier output level at 10 μs. It should be completely settled and unchanged from the 1 μs value. Any significant magnitude variations must be resolved before proceeding to measure OUT settling time. (If a significant change between 1 and 10 μs is noticed, it is probably due to a thermal unbalance in the amplifier fixture rather than a problem with the OUT. To check this, perform a t8HL measurement between 1 and 10 μs. If the change is similar but of opposite polarity, it can be attributed to the fixture and ignored.)
FIGURE 3. Test circuit for settling time - Continued.