MIL-M-38510/7588
TA8LE I. Electrical performance characteristics - Continued.
Test and MIL-STD-883 test method |
Symbol |
Test conditions 1/ -55°C ≤ TC ≤ +125°C 3.0 V ≤ VCC ≤ 5.5 V unless otherwise specified |
Device type 2/ |
VCC |
Group A subgroups |
Limits 1/ |
Unit |
||
Min |
Max |
||||||||
Propagation delay time, E3 to On 3003 |
tPHL3, tPLH3 4/ 5/ 10/11/ |
CL = 50 pF RL = 5000 |
02 |
3.0 V |
9, 11 |
1.0 |
15.5 |
ns |
|
10 |
1.0 |
17.0 |
|||||||
see figure 3 |
M |
02 |
9 |
1.0 |
15.5 |
||||
D |
02 |
1.0 |
15.5 |
||||||
P, L, R |
02 |
1.0 |
15.5 |
||||||
CL = 50 pF RL = 5000 |
02 |
4.5 V |
9, 11 |
1.0 |
11.5 |
ns |
|||
10 |
1.0 |
13.5 |
|||||||
see figure 3 |
M |
02 |
9 |
1.0 |
11.5 |
||||
D |
02 |
1.0 |
11.5 |
||||||
P, L, R |
02 |
1.0 |
11.5 |
1/ Each input/output, as applicable, shall be tested at the specified temperature for the specified limits. Output terminals not designated shall be high level logic, low level logic, or open, except as follows:
a. VIC (pos) tests, the GND terminal can be open. TC = +25°C. b. VIC (neg) tests, the VCC terminal shall be open. TC = +25°C.
c. All ICC tests, the output terminal shall be open. When performing these tests, the current meter shall be placed in the circuit such that all current flows through the meter.
For negative and positive voltage and current values, the sign designates the potential difference in reference to GND and the direction of current flow, respectively; and the absolute value of the magnitude, not the sign, is relative to the minimum and maximum limits, as applicable, listed herein.
2/ The word "All" in the device type column means non-RHA limits for all devices types. M, D, P, L, and R in the conditions column specify the postirradiation limits for those device types specified in the device type column.
3/ This test is guaranteed, if not tested, to the limits specified in table I.
4/ RHA samples do not have to be tested at -55°C and +125°C prior to irradiation.
5/ When performing postirradiation electrical measurements for any RHA level, TA = +25°C. Limits shown are guaranteed at
TA = +25°C ±5°C.
6/ Transmission driving tests are performed at VCC = 5.5 V dc with a 2 ms duration maximum.
7/ Power dissipation capacitance (CPD) determines the no load dynamic power consumption,
PD = (CPD + CL) (VCC x VCC)f + (ICC x VCC) and the dynamic current consumption, IS = (CPD + CL)VCCf + ICC. For both CPD and
IS, f is the frequency of the input signal.
8/ See EIA/JEDEC STD. No. 78 for electrically induced latch-up test methods and procedures. The values listed for Itrigger and
Vover are to be accurate within ±5 percent.
9/ Tests shall be performed in sequence, attributes data only. Functional tests shall include the truth tables and other logic patterns used for fault detection. Functional tests shall be performed in sequence as approved by the qualifying activity on qualified devices. H � 2.5 V, L < 2.5 V; high inputs = 3.7 V and low inputs = 0.6 V for VCC = 4.5 V and H � 1.5 V, L < 1.5 V; high inputs = 2.5 V and low inputs = 0.45 V for VCC = 3.0 V. Tests at VCC = 3.0 V are for RHA specified devices only
(TA = +25°C ±5°C). Functional tests at VCC = 3.0 V are worst case for RHA specified devices.
10/ Devices are tested at VCC = 3.0 V and VCC = 4.5 V at TC = +125°C for sample testing and at VCC = 3.0 V and VCC = 4.5 V at
TC = +25°C for screening. Other voltages of VCC and temperatures are guaranteed, if not tested. See 4.4.1d.
11/ AC limits at VCC = 5.5 V are equal to the limits at VCC = 4.5 V and guaranteed by testing at VCC = 4.5 V. Minimum ac limits for VCC = 5.5 V are 1.0 ns and guaranteed by guardbanding the VCC = 4.5 V minimum limits to 1.5 ns. For propagation delay tests, all paths must be tested.
10
For Parts Inquires call Parts Hangar, Inc (727) 493-0744
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business